For absolute pressure (from the differential pressure series) Siemens

Чертеж
SITRANS P pressure transmitters, DS III HART series for absolute pressure, from the differential pressure series, dimensions in mm (inch)
SITRANS P pressure transmitters, DS III PA and FF series for absolute pressure, from the differential pressure series, dimensions in mm (inch)
Технические данные
SITRANS P, DS III series for absolute pressure (from the differential pressure series) | ||||
---|---|---|---|---|
HART | PROFIBUS PA or Foundation Fieldbus | |||
Input | ||||
Measured variable | Absolute pressure | |||
Spans (infinitely adjustable) or | Span | Maximum operating pressure | Nominal measuring range | Maximum operating pressure |
8.3 ... 250 mbar a | 32 bar a | 250 mbar a | 32 bar a | |
43 ... 1300 mbar a | 32 bar a | 1300 bar a | 32 bar a | |
160 ... 5000 mbar a | 32 bar a | 5 bar a | 32 bar a | |
1 ... 30 bar a | 160 bar a | 30 bar a | 160 bar a | |
5.3 ... 100 bar a | 160 bar a | 100 bar a | 160 bar a | |
Lower measuring limit | ||||
| 0 mbar a (0 psi a) | |||
Upper measuring limit | 100% of max. span | |||
Output | ||||
Output signal | 4 ... 20 mA | Digital PROFIBUS PA or Foundation Fieldbus signal | ||
| 3.55 mA, factory preset to 3.84 mA | - | ||
| 23 mA, factory preset to 20.5 mA or optionally set to 22.0 mA | - | ||
Load | ||||
| RB ≤ (UH - 10.5 V)/0.023 A in Ω, | - | ||
| RB = 230 ... 500 Ω (SIMATIC PDM) or | - | ||
Physical bus | - | IEC 61158-2 | ||
Protection against polarity reversal | Protected against short-circuit and polarity reversal. Each connection against the other with max. supply voltage. | |||
Measuring accuracy | Acc. to EN 60770-1 | |||
Reference conditions | Increasing characteristic, start-of-scale value 0 bar, stainless steel seal diaphragm, silicone oil filling, room temperature 25 °C (77 °F)) r: Span ratio | |||
Error in measurement and fixed-point setting (including hysteresis and repeatability) | ||||
| ≤ 0,1 % | |||
| ≤ 0,1 % | |||
| ≤ 0,2 % | |||
Long-term drift (temperature change ±30 °C (±54 °F)) | ≤ (0.1 · r)%/year | ≤ 0.1 %/year | ||
Influence of ambient temperature | ||||
| ≤ (0.1 · r + 0.2)% | ≤ 0,3 % | ||
| ≤ (0.1·r + 0.15) %/10 K | ≤ 0.25 %/10 K | ||
Measured Value Resolution | - | 3 · 10-5 of nominal measuring range | ||
Rated conditions | ||||
Degree of protection (to EN 60529) | IP65 | |||
Temperature of medium | ||||
| -40 ... +100 °C (-40 ... +212 °F) | |||
| -20 ... +100 °C (-4 ... +212 °F) | |||
| -20 ... +60 °C (-4 ... +140 °F) | |||
Ambient conditions | ||||
| ||||
| -30 ... +85 °C (-22 ... +185 °F) | |||
| -50 ... +85 °C (-58 ... +185 °F) | |||
| ||||
| Permissible | |||
| ||||
| Acc. to EN 50081-1 | |||
| Acc. to EN 61326 and NAMUR NE 21 | |||
Design | ||||
Weight (without options) | ≈ 4.5 kg (≈ 9.9 lb) | |||
Enclosure material | Low-copper die-cast aluminum, GD-AlSi 12 or stainless steel precision casting, mat. No. 1.4408 | |||
Wetted parts materials | ||||
| Stainless steel, mat. No. 1.4404/316L or Hastelloy C276, mat. No. 2.4819, Monel, mat. No. 2.4360, tantalum or gold | |||
| Stainless steel, mat. No. 1.4408, Hastelloy C4, mat. No. 2.4610 or Monel, mat. No. 2.4360 | |||
| FPM (Viton) or optionally: PTFE, FEP, FEPM and NBR | |||
Measuring cell filling | Silicone oil or inert filling liquid (max. 160 bar (2320 psi a) with oxygen measurement) | |||
Process connection | ¼-18 NPT and flange connection with mounting thread M10 to DIN 19213 or 7/16-20 UNF to EN 61518 | |||
Material of mounting bracket | ||||
| Sheet-steel, Mat. No. 1.0330, chrome-plated | |||
| Sheet stainless steel, mat. No. 1.4301 (SS 304) | |||
Power supply UH | Supplied through bus | |||
Terminal voltage on transmitter | 10.5 ... 45 V DC | - | ||
Separate 24 V power supply necessary | - | No | ||
Bus voltage | ||||
| - | 9 ...32 V | ||
| - | 9 ...24 V | ||
Current consumption | ||||
| - | 12.5 mA | ||
| - | Yes | ||
| - | 15.5 mA | ||
Fault disconnection electronics (FDE) available | - | Yes | ||
Certificates and approvals | ||||
Classification according to PED 97/23/EC | For gases of fluid group 1 and liquids of fluid group 1; complies with requirements of article 3, paragraph 3 (sound engineering practice) | |||
Explosion protection | ||||
| PTB 99 ATEX 2122 | |||
| Ex II 1/2 G EEx ia/ib IIB/IIC T6 | |||
| -40 ... +85 °C (-40 ... +185 °F) temperature class T4; | |||
| To certified intrinsically-safe circuits with peak values: | FISCO supply unit: Linear barrier: | ||
| Li = 0.4 mH, Ci = 6 nF | Li = 7 μH, Ci = 1.1 nF | ||
| PTB 99 ATEX 1160 | |||
| Ex II 1/2 G EEx d IIC T4/T6 | |||
| -40 ... +85 °C (-40 ... +185 °F) temperature class T4; | |||
| To circuits with values: UH = 10.5 ... 45 V DC | To circuits with values: UH = 9 ... 32 V DC | ||
| PTB 01 ATEX 2055 | |||
| Ex II 1 D IP65 T 120 °C | |||
| -40 ... +85 °C (-40 ... +185 °F) | |||
| 120 °C (248 °F) | |||
| To certified intrinsically-safe circuits with peak values: | FISCO supply unit: Linear barrier: | ||
| Li = 0.4 mH, Ci = 6 nF | Li = 7 μH, Ci = 1.1 nF | ||
| PTB 01 ATEX 2055 | |||
| Ex II 2 D IP65 T 120 °C | |||
| To circuits with values: UH = 10.5 ... 45 V DC; Pmax = 1.2 W | To circuits with values: UH = 9 ... 32 V DC; Pmax = 1.2 W | ||
| TÜV 01 ATEX 1696 X | Planned | ||
| Ex II 3 G EEx nA L IIC T4/T5/T6 | - | ||
| Certificate of Compliance 3008490 | |||
| CL I, DIV 1, GP ABCD T4...T6; CL II, DIV 1, GP EFG; CL III; CL I, ZN 0/1 AEx ia IIC T4...T6; CL I, DIV 2, GP ABCD T4...T6; CL II, DIV 2, GP FG; CL III | |||
| Certificate of Compliance 1153651 | |||
| CL I, DIV 1, GP ABCD T4...T6; CL II, DIV 1, GP EFG; CL III; Ex ia IIC T4...T6; CL I, DIV 2, GP ABCD T4...T6; CL II, DIV 2, GP FG; CL III |
HART communication | |
---|---|
HART communication | 230 ... 1100 Ω |
Protocol | HART Version 5.x |
Software for computer | SIMATIC PDM |
PROFIBUS PA communication | |
---|---|
Simultaneous communication with master class 2 (max.) | 4 |
The address can be set using | Configuration tool or local operation (standard setting address 126) |
Cyclic data usage | |
| 5 (one measured value) or |
| 0, 1, or 2 (register operating mode and reset function for metering) |
Internal preprocessing | |
Device profile | PROFIBUS PA Profile for Process Control Devices Version 3.0, Class B |
Function blocks | 2 |
| |
| Yes, linearly rising or falling characteristic |
| 0 to 100 s |
| Input /Output |
| parameterizable (last good value, substitute value, incorrect value) |
| Yes, one upper and lower warning limit and one alarm limit respectively |
| Can be reset, preset, optional direction of counting, simulation function of register output |
| parameterizable (summation with last good value, continuous summation, summation with incorrect value) |
| One upper and lower warning limit and one alarm limit respectively |
| 1 |
Transducer blocks | 2 |
| |
| Yes |
| Yes |
| Max. 30 nodes |
| Yes |
| Parameterizable |
| Constant value or over parameterizable ramp function |
Communication Foundation Fieldbus | |
---|---|
Function blocks | 3 function blocks analog input, 1 function block PID |
| |
| Yes, linearly rising or falling characteristic |
| 0 to 100 s |
| Output/input (can be locked within the device with a bridge) |
| parameterizable (last good value, substitute value, incorrect value) |
| Yes, one upper and lower warning limit and one alarm limit respectively |
| Yes |
| Standard FF function block |
| 1 resource block |
Transducer blocks | 1 transducer block Pressure with calibration, 1 transducer block LCD |
| |
| Yes |
| Yes |
| Constant value or over parameterizable ramp function |