19 rack unit and portable version Siemens

Чертеж



ULTRAMAT 23, 19“ unit, dimensions in mm

ULTRAMAT 23, desktop unit, dimensions in mm

Схема подключения



Pin assignment (electrical and gas connections)

ULTRAMAT 23, pin assignment (standard)

ULTRAMAT 23, pin assignment of the optional PROFIBUS interface board

19“ unit

ULTRAMAT 23, 19“ unit, e.g. one infrared component with oxygen measurement

Desktop unit

ULTRAMAT 23, portable unit, in sheet-steel housing, gas and electrical connections

ULTRAMAT 23, designation of the different labels

Технические данные



ULTRAMAT 23, 19" rack unit and portable version

General information

Measured components

Maximum of 4, comprising three infrared-sensitive gases and oxygen

Measuring ranges

Two per measured component

Display

LCD with LED backlighting and contrast control; function keys; 80 characters (4 lines/20 characters)

Operating position

Front wall, vertical

Conformity

CE symbol EN 61000-6-2, EN 61000-6-4

Design, enclosure

Weight

Approximately 10 kg

Degree of protection, 19" rack unit and desktop model

IP20 according to EN 60529

Electrical characteristics

EMC (Electromagnetic Compatibility) (safety extra-low voltage (SELV) with safety isolation)

In accordance with standard requirements of NAMUR NE21 (08/98) or EN 50081-1, EN 50082-2

Auxiliary power

100 V AC, +10%/-15%, 50 Hz,

120 V AC, +10%/-15%, 50 Hz,

200 V AC, +10%/-15%, 50 Hz,

230 V AC, +10%/-15%, 50 Hz,

100 V AC, +10%/-15%, 60 Hz,

120 V AC, +10%/-15%, 60 Hz,

230 V AC, +10%/-15%, 60 Hz

Power consumption

Approx. 60 VA

Electrical inputs and outputs

Analog output

Per component, 0/2/4 up to 20 mA, NAMUR, isolated, max. load 750 Ω

Relay outputs

8, with changeover contacts, freely parameterizable, e.g. for measuring range identification;

24 V AC/DC/1 A load, potential-free, non-sparking

Binary inputs

3, dimensioned for 24 V, potential-free

 
  • Pump
 
  • AUTOCAL
 
  • Synchronization

Serial interface

RS 485

AUTOCAL function

Automatic unit calibration with ambient air (depending on measured component); adjustable cycle time from 0 (1) …  24 hours

Options

Add-on electronics, each with 8 additional digital inputs and relay outputs for e.g. triggering of automatic calibration and for PROFIBUS PA or PROFIBUS DP

Climatic conditions

Permissible ambient temperature

 
  • During operation

5 … 45 °C

  • During storage and transportation

-20 ... +60 °C

Permissible ambient humidity

< 90% RH (relative humidity) during storage and transportation

Permissible pressure fluctuations

600 … 1 200 hPa

Gas inlet conditions

Sample gas pressure

 
  • Without pump

Unpressurized (< 1 200 hPa, absolute)

  • With pump

Depressurized suction mode, set in factory with 2 m hose at sample gas outlet; full-scale value calibration necessary under different venting conditions
(800 ... 1 050 hPa, absolute)

Sample gas flow

72 … 120 l/h (1.2 … 2 l/min)

Sample gas temperature

Min. 0 to max. 50 °C, but above the dew point

Sample gas humidity

< 90% RH (relative humidity), non-condensing



Technical data, infrared channel

So that the technical data can be complied with, a cycle time of ≤ 24 hours must be activated for the AUTOCAL. The cycle time of the AUTOCAL function must be ≤ 6 hours when measuring small NO and SO2 measuring ranges (≤ 400 mg/m³) on TÜV/QAL-certified systems.

Measuring ranges

See ordering data

Chopper compartment flushing

Upstream pressure approximately 3 000 hPa; purging gas consumption approximately 100 ml/min

Time response

Warm-up period

Approximately 30 min (at room temperature) (the technical specification will be met after 2 hours)

Delayed display (T90 time)

Dependent on length of analyzer chamber, sample gas line and parameterizable attenuation

Damping (electrical time constant)

Parameterizable from 0 … 99.9 s

Measuring response

(relating to sample gas pressure 1 013 hPa absolute, 1.0 l/min sample gas flow and 25 °C ambient temperature)

Output signal fluctuation

< ± 1% of the current measuring range (see rating plate)

Detection limit

1% of the current measuring range

Linearity error

  • In largest possible measuring range:
    < ± 1% of the full-scale value
  • In smallest possible measuring range:
    < ± 2% of the full-scale value

Repeatability

≤ ± 1% of the current measuring range

Drift

Zero point

 
  • With AUTOCAL

Negligible

  • Without AUTOCAL

< 2 % of the current measuring range/week

Full-scale value drift

 
  • With AUTOCAL

Negligible

  • Without AUTOCAL

< 2 % of the current measuring range/week

Influencing variables

(relating to sample gas pressure 1 013 hPa absolute, 1.0 l/min sample gas flow and 25 °C ambient temperature)

Temperature

Max. 2% of the smallest possible measuring range according to rating plate per 10 K with an AUTOCAL cycle time of 6 h

Atmospheric pressure

< 0.2% of the current measuring range per 1% pressure variation

Auxiliary power

< 0.1% of the current measuring range with a change of ± 10%



Technical data, oxygen channel (electrochemical)

Measuring ranges

0 … 5 % … 0 … 25 % O2, parameterizable

Service life

Approximately 2 years at 21% O2; continuous duty < 0.5% O2 will destroy the measuring cell

Detection limit

1% of the current measuring range

Time response

Delayed display (T90 time)

Dependent on dead time and parameterizable attenuation, not > 30 s at approximately 1.2 l/min sample gas flow

Measuring response

(relating to sample gas pressure 1 013 hPa absolute, 1.0 l/min sample gas flow and 25 °C ambient temperature)

Output signal fluctuation

< ± 0.5% of the current measuring range

Linearity error

< ± 0.2% of the current measuring range

Repeatability

≤ 0.05 % O2

Drift

 
  • With AUTOCAL

Negligible

  • Without AUTOCAL

1% O2/year in air, typical

Influencing variables

(relating to sample gas pressure 1 013 hPa absolute, 1.0 l/min sample gas flow and 25 °C ambient temperature)

Temperature

< ± 0.5% O2 per 20 K, relating to a measured value at 20 °C

Atmospheric pressure

< 0.2% of the measured value per 1% pressure variation

Accompanying gases

The oxygen sensor must not be used if the accompanying gas contains the following components: Chlorine or fluorine compounds, heavy metals, aerosols, mercaptans, alkaline components (such as NH3 in % range)

Typical combustion exhaust gases

Influence: < 0.05% O2

Humidity

H2O dew point ≥ 2 °C; the oxygen sensor must not be used with dry sample gases (however, no condensation either)



Technical data, H2S channel for measuring ranges of 5 ... 50 vpm

Measured components

Maximum of four, comprising one or two infrared-sensitive gases, one oxygen component and one hydrogen sulfide component

Measuring ranges

 
  • Smallest measuring range

0 ... 5 vpm

  • Largest measuring range

0 ... 50 vpm

Service life of the sensor

Approx. 12 months

Permissible atmospheric pressure

750 ... 1 200 hPa

Permissible operating temperature

5 ... 40 °C (41 ... 104 °F)

Operating mode

Continuous measurement between 0 and 12.5 vpm
Discontinuous measurement between 12.5 and 50 vpm

Influencing variables

Accompanying gases

The hydrogen sulfide sensor must not be used if the accompanying gas contains the following components:

  • Compounds containing chlorine
  • Compounds containing fluorine
  • Heavy metals
  • Aerosols
  • Alkaline components
    (e.g. NH3 > 5%)

Cross-inferences (interfering gases)

1 360 vpm SO2 result in a cross-interference of < 20 vpm H2S

180 vpm NO result in a cross-interference of < 150 vpm H2S

No cross-interference of CH4, CO2 and H2 (1 000 vpm)

Drift

< 1% of the current measuring range per month

Temperature

< 3% /10 K referred to full-scale value

Atmospheric pressure

< 0.2% of the measured value per 1% pressure variation

Measuring response

Delayed display (T90 time)

< 40 s with sample gas flow of approx. 1 ... 1.2 l/min

Output signal noise

< 2% of smallest measuring range with an attenuation constant of 30 s

Display resolution

< 0.01 vpm H2S

Output signal resolution

< 1% of smallest measuring range with an attenuation constant of 30 s

Repeatability

< 4% of smallest measuring range



Technical data, H2S channel for measuring ranges of 0 ... 5 000 vpm

Measured components

Maximum of four, comprising one or two infrared-sensitive gases, one oxygen component and one hydrogen sulfide component

Measuring ranges of H2S sensor MB 5000

 
  • Smallest measuring range

0 ... 5 000 vpm

  • Largest measuring range

0 ... 5 000 vpm

Service life of the sensor

Approx. 12 months

Permissible atmospheric pressure

750 ... 1 200 hPa

Permissible operating temperature

5 ... 40 °C (41 ... 104 °F)

Influencing variables

Accompanying gases

The hydrogen sulfide sensor must not be used if the accompanying gas contains the following components:

  • Compounds containing chlorine
  • Compounds containing fluorine
  • Heavy metals
  • Aerosols
  • Alkaline components
    (e.g. NH3 > 5 vpm)

Cross-inferences (interfering gases)

100 ppm SO2 result in a cross-interference of < 30 ppm H2S

Drift

< 1% of the current measuring range per month

Temperature

< 3% /10 K referred to full-scale value

Atmospheric pressure

< 0.2% of the measured value per 1% pressure variation

Measuring response

Delayed display (T90 time)

< 80 s with sample gas flow of approx. 1 ... 1.2 l/min

Output signal noise

< 15 ppm H2S

Display resolution

< 0.2% of the full-scale value

Output signal resolution

< 30 ppm H2S

Repeatability

< 4% referred to full-scale value



Technical data, paramagnetic oxygen cell

Measured components

Maximum of four, comprising up to three infrared-sensitive gases and an oxygen component

Measuring ranges

Two per component

  • Min. 0 ... 2% vol O2
  • Max. 0 ... 100% vol O2
  • Suppressed measuring range possible; e.g. 95 ... 100%

Permissible atmospheric pressure

700 ... 1 200 hPa

Permissible operating temperature

5 ... 45 °C (41 ... 113 °F)

Cross-inferences (interfering gases)

See Table "Cross-sensitivities"

Zero point drift

  • Measuring range 2%: max. 0.1% with weekly zero adjustment
  • Measuring range 5%: max. 0.1% with weekly zero adjustment
  • Measuring range 25% or greater:
    max. 0.5% with monthly zero adjustment

Measured-value drift

Negligible with AUTOCAL

Temperature error

< 2%/10 K referred to measuring range 5%

< 5%/10 K referred to measuring range 2%

Humidity error for N2 with 90% relative humidity after 30 min

< 0.6% at 50 °C

Atmospheric pressure

< 0.2% of measured value per 1% pressure variation

Delayed display (T90 time)

< 60 s

Output signal noise

< 1% of smallest measuring range

Repeatability

< 1% of smallest measuring range



Дальнейшая информация



Ordering notes

Special selection rules must be observed when measuring some components.

Measured component N2O

7MB2335, 7MB2337 and 7MB2338
(application: Si chip production)

  • Measuring range 0 to 100 / 500 ppm (MB designation "E")
  • Can only be used to measure N2O in ultra-pure gases

7MB2337 and 7MB2338
(application: measurement in accordance with the requirements of the Kyoto protocol)

  • Measuring range 0 to 500 / 5 000 vpm (MB designation "Y")
  • Requires simultaneous measurement of CO2 for correction of cross-interference

7MB2337-*CP*0-*SY* or
7MB2338-*DC*0-*SY* (including NO measurement)

7MB2337 and 7MB2338
(application with paramagnetic oxygen measuring cell and separate gas path)

7MB2337-4**80-**** - Z + C11
7MB2337-5**80-**** - Z + C11

7MB2338-4**80-**** - Z + C11
7MB2338-5**80-**** - Z + C11

Measured component SF6

7MB2335, 7MB2337 and 7MB2338
(application: SI chip production)

  • Measuring range 0 to 500 / 2 500 ppm (MB designation "H")
  • Can only be used to measure SF6 in inert gases

Calibration interval (MCERTs versions 7MB2335, 7MB2337, 7MB2338)

Component

Smallest measuring range

Calibration interval

Remarks

Z suffix

CO

0 … 150 mg/m³

5 months

IED 2010/75/EC

E50

CO

0 … 250 mg/m³

12 months

IED 2010/75/EC

 

NO

0 … 100 mg/m³

5 months

IED 2010/75/EC

E50

NO

0 … 250 mg/m³

12 months

IED 2010/75/EC

 

SO2

0 … 400 mg/m³

12 months

IED 2010/75/EC

 

N2O

0 … 500 ppm

 

Kyoto protocol

 

N2O

0 … 50 mg/m³

6 months

IED 2010/75/EC

 


Calibration interval (TÜV versions 7MB2355, 7MB2357, 7MB2358)

Component

Smallest measuring range

Calibration interval

Remarks

Z suffix

CO

0 ... 200 mg/m³

1 month

13th/27th BImSchV

TX3

NO

0 ... 150 mg/m³

1 month

13th/27th BImSchV

TX3

SO2

0 ... 400 mg/m³

1 month

13th/27th BImSchV

TX3



 

AUTOCAL
(ambient air)

AUTOCAL
(inert gas e.g. N2)

 

Calibration with calibration gas

Comment (keep to technical specs)

Zero point

Calibration point

Zero point

Calibration point

Zero point

Calibration point

Hours

Weeks

  

IR components

3 ... 24

3 ... 24

o

52

 

O2 - electrical chemical sensor

Stable

3 ... 24

Stable

-

52

o

 

O2 paramagnetic Cell

-

3 ... 24

x

x

1

o

at MB < 5%

-

3 ... 24

x

x

8

o

at MB > 5%

O2 paramagnetic Cell

x

x

3 ... 24

-

o

52

at MB < 5%

x

x

3 ... 24

-

o

52

at MB > 5%

H2S sensor

3

-

3

-

o

12

 

o = with AUTOCAL, x = not applicable



Calibration intervals, standard devices