19 rack unit and portable version Siemens

Чертеж
ULTRAMAT 23, 19“ unit, dimensions in mm
ULTRAMAT 23, desktop unit, dimensions in mm
Схема подключения
Pin assignment (electrical and gas connections)
ULTRAMAT 23, pin assignment (standard)
ULTRAMAT 23, pin assignment of the optional PROFIBUS interface board
19“ unit
ULTRAMAT 23, 19“ unit, e.g. one infrared component with oxygen measurement
Desktop unit
ULTRAMAT 23, portable unit, in sheet-steel housing, gas and electrical connections
ULTRAMAT 23, designation of the different labels
Технические данные
ULTRAMAT 23, 19" rack unit and portable version | |
General information | |
Measured components | Maximum of 4, comprising three infrared-sensitive gases and oxygen |
Measuring ranges | Two per measured component |
Display | LCD with LED backlighting and contrast control; function keys; 80 characters (4 lines/20 characters) |
Operating position | Front wall, vertical |
Conformity | CE symbol EN 61000-6-2, EN 61000-6-4 |
Design, enclosure | |
Weight | Approximately 10 kg |
Degree of protection, 19" rack unit and desktop model | IP20 according to EN 60529 |
Electrical characteristics | |
EMC (Electromagnetic Compatibility) (safety extra-low voltage (SELV) with safety isolation) | In accordance with standard requirements of NAMUR NE21 (08/98) or EN 50081-1, EN 50082-2 |
Auxiliary power | 100 V AC, +10%/-15%, 50 Hz, 120 V AC, +10%/-15%, 50 Hz, 200 V AC, +10%/-15%, 50 Hz, 230 V AC, +10%/-15%, 50 Hz, 100 V AC, +10%/-15%, 60 Hz, 120 V AC, +10%/-15%, 60 Hz, 230 V AC, +10%/-15%, 60 Hz |
Power consumption | Approx. 60 VA |
Electrical inputs and outputs | |
Analog output | Per component, 0/2/4 up to 20 mA, NAMUR, isolated, max. load 750 Ω |
Relay outputs | 8, with changeover contacts, freely parameterizable, e.g. for measuring range identification; 24 V AC/DC/1 A load, potential-free, non-sparking |
Binary inputs | 3, dimensioned for 24 V, potential-free |
| |
| |
| |
Serial interface | RS 485 |
AUTOCAL function | Automatic unit calibration with ambient air (depending on measured component); adjustable cycle time from 0 (1) … 24 hours |
Options | Add-on electronics, each with 8 additional digital inputs and relay outputs for e.g. triggering of automatic calibration and for PROFIBUS PA or PROFIBUS DP |
Climatic conditions | |
Permissible ambient temperature | |
| 5 … 45 °C |
| -20 ... +60 °C |
Permissible ambient humidity | < 90% RH (relative humidity) during storage and transportation |
Permissible pressure fluctuations | 600 … 1 200 hPa |
Gas inlet conditions | |
Sample gas pressure | |
| Unpressurized (< 1 200 hPa, absolute) |
| Depressurized suction mode, set in factory with 2 m hose at sample gas outlet; full-scale value calibration necessary under different venting conditions |
Sample gas flow | 72 … 120 l/h (1.2 … 2 l/min) |
Sample gas temperature | Min. 0 to max. 50 °C, but above the dew point |
Sample gas humidity | < 90% RH (relative humidity), non-condensing |
Technical data, infrared channel | |
So that the technical data can be complied with, a cycle time of ≤ 24 hours must be activated for the AUTOCAL. The cycle time of the AUTOCAL function must be ≤ 6 hours when measuring small NO and SO2 measuring ranges (≤ 400 mg/m³) on TÜV/QAL-certified systems. | |
Measuring ranges | See ordering data |
Chopper compartment flushing | Upstream pressure approximately 3 000 hPa; purging gas consumption approximately 100 ml/min |
Time response | |
Warm-up period | Approximately 30 min (at room temperature) (the technical specification will be met after 2 hours) |
Delayed display (T90 time) | Dependent on length of analyzer chamber, sample gas line and parameterizable attenuation |
Damping (electrical time constant) | Parameterizable from 0 … 99.9 s |
Measuring response (relating to sample gas pressure 1 013 hPa absolute, 1.0 l/min sample gas flow and 25 °C ambient temperature) | |
Output signal fluctuation | < ± 1% of the current measuring range (see rating plate) |
Detection limit | 1% of the current measuring range |
Linearity error |
|
Repeatability | ≤ ± 1% of the current measuring range |
Drift | |
Zero point | |
| Negligible |
| < 2 % of the current measuring range/week |
Full-scale value drift | |
| Negligible |
| < 2 % of the current measuring range/week |
Influencing variables (relating to sample gas pressure 1 013 hPa absolute, 1.0 l/min sample gas flow and 25 °C ambient temperature) | |
Temperature | Max. 2% of the smallest possible measuring range according to rating plate per 10 K with an AUTOCAL cycle time of 6 h |
Atmospheric pressure | < 0.2% of the current measuring range per 1% pressure variation |
Auxiliary power | < 0.1% of the current measuring range with a change of ± 10% |
Technical data, oxygen channel (electrochemical) | |
Measuring ranges | 0 … 5 % … 0 … 25 % O2, parameterizable |
Service life | Approximately 2 years at 21% O2; continuous duty < 0.5% O2 will destroy the measuring cell |
Detection limit | 1% of the current measuring range |
Time response | |
Delayed display (T90 time) | Dependent on dead time and parameterizable attenuation, not > 30 s at approximately 1.2 l/min sample gas flow |
Measuring response (relating to sample gas pressure 1 013 hPa absolute, 1.0 l/min sample gas flow and 25 °C ambient temperature) | |
Output signal fluctuation | < ± 0.5% of the current measuring range |
Linearity error | < ± 0.2% of the current measuring range |
Repeatability | ≤ 0.05 % O2 |
Drift | |
| Negligible |
| 1% O2/year in air, typical |
Influencing variables (relating to sample gas pressure 1 013 hPa absolute, 1.0 l/min sample gas flow and 25 °C ambient temperature) | |
Temperature | < ± 0.5% O2 per 20 K, relating to a measured value at 20 °C |
Atmospheric pressure | < 0.2% of the measured value per 1% pressure variation |
Accompanying gases | The oxygen sensor must not be used if the accompanying gas contains the following components: Chlorine or fluorine compounds, heavy metals, aerosols, mercaptans, alkaline components (such as NH3 in % range) |
Typical combustion exhaust gases | Influence: < 0.05% O2 |
Humidity | H2O dew point ≥ 2 °C; the oxygen sensor must not be used with dry sample gases (however, no condensation either) |
Technical data, H2S channel for measuring ranges of 5 ... 50 vpm | |
Measured components | Maximum of four, comprising one or two infrared-sensitive gases, one oxygen component and one hydrogen sulfide component |
Measuring ranges | |
| 0 ... 5 vpm |
| 0 ... 50 vpm |
Service life of the sensor | Approx. 12 months |
Permissible atmospheric pressure | 750 ... 1 200 hPa |
Permissible operating temperature | 5 ... 40 °C (41 ... 104 °F) |
Operating mode | Continuous measurement between 0 and 12.5 vpm |
Influencing variables | |
Accompanying gases | The hydrogen sulfide sensor must not be used if the accompanying gas contains the following components:
|
Cross-inferences (interfering gases) | 1 360 vpm SO2 result in a cross-interference of < 20 vpm H2S 180 vpm NO result in a cross-interference of < 150 vpm H2S No cross-interference of CH4, CO2 and H2 (1 000 vpm) |
Drift | < 1% of the current measuring range per month |
Temperature | < 3% /10 K referred to full-scale value |
Atmospheric pressure | < 0.2% of the measured value per 1% pressure variation |
Measuring response | |
Delayed display (T90 time) | < 40 s with sample gas flow of approx. 1 ... 1.2 l/min |
Output signal noise | < 2% of smallest measuring range with an attenuation constant of 30 s |
Display resolution | < 0.01 vpm H2S |
Output signal resolution | < 1% of smallest measuring range with an attenuation constant of 30 s |
Repeatability | < 4% of smallest measuring range |
Technical data, H2S channel for measuring ranges of 0 ... 5 000 vpm | |
Measured components | Maximum of four, comprising one or two infrared-sensitive gases, one oxygen component and one hydrogen sulfide component |
Measuring ranges of H2S sensor MB 5000 | |
| 0 ... 5 000 vpm |
| 0 ... 5 000 vpm |
Service life of the sensor | Approx. 12 months |
Permissible atmospheric pressure | 750 ... 1 200 hPa |
Permissible operating temperature | 5 ... 40 °C (41 ... 104 °F) |
Influencing variables | |
Accompanying gases | The hydrogen sulfide sensor must not be used if the accompanying gas contains the following components:
|
Cross-inferences (interfering gases) | 100 ppm SO2 result in a cross-interference of < 30 ppm H2S |
Drift | < 1% of the current measuring range per month |
Temperature | < 3% /10 K referred to full-scale value |
Atmospheric pressure | < 0.2% of the measured value per 1% pressure variation |
Measuring response | |
Delayed display (T90 time) | < 80 s with sample gas flow of approx. 1 ... 1.2 l/min |
Output signal noise | < 15 ppm H2S |
Display resolution | < 0.2% of the full-scale value |
Output signal resolution | < 30 ppm H2S |
Repeatability | < 4% referred to full-scale value |
Technical data, paramagnetic oxygen cell | |
Measured components | Maximum of four, comprising up to three infrared-sensitive gases and an oxygen component |
Measuring ranges | Two per component
|
Permissible atmospheric pressure | 700 ... 1 200 hPa |
Permissible operating temperature | 5 ... 45 °C (41 ... 113 °F) |
Cross-inferences (interfering gases) | See Table "Cross-sensitivities" |
Zero point drift |
|
Measured-value drift | Negligible with AUTOCAL |
Temperature error | < 2%/10 K referred to measuring range 5% < 5%/10 K referred to measuring range 2% |
Humidity error for N2 with 90% relative humidity after 30 min | < 0.6% at 50 °C |
Atmospheric pressure | < 0.2% of measured value per 1% pressure variation |
Delayed display (T90 time) | < 60 s |
Output signal noise | < 1% of smallest measuring range |
Repeatability | < 1% of smallest measuring range |
Дальнейшая информация
Ordering notes
Special selection rules must be observed when measuring some components.
Measured component N2O
7MB2335, 7MB2337 and 7MB2338
(application: Si chip production)
- Measuring range 0 to 100 / 500 ppm (MB designation "E")
- Can only be used to measure N2O in ultra-pure gases
7MB2337 and 7MB2338
(application: measurement in accordance with the requirements of the Kyoto protocol)
- Measuring range 0 to 500 / 5 000 vpm (MB designation "Y")
- Requires simultaneous measurement of CO2 for correction of cross-interference
7MB2337-*CP*0-*SY* or
7MB2338-*DC*0-*SY* (including NO measurement)
7MB2337 and 7MB2338
(application with paramagnetic oxygen measuring cell and separate gas path)
7MB2337-4**80-**** - Z + C11
7MB2337-5**80-**** - Z + C11
7MB2338-4**80-**** - Z + C11
7MB2338-5**80-**** - Z + C11
Measured component SF6
7MB2335, 7MB2337 and 7MB2338
(application: SI chip production)
- Measuring range 0 to 500 / 2 500 ppm (MB designation "H")
- Can only be used to measure SF6 in inert gases
Calibration interval (MCERTs versions 7MB2335, 7MB2337, 7MB2338)
Component | Smallest measuring range | Calibration interval | Remarks | Z suffix |
---|---|---|---|---|
CO | 0 … 150 mg/m³ | 5 months | IED 2010/75/EC | E50 |
CO | 0 … 250 mg/m³ | 12 months | IED 2010/75/EC | |
NO | 0 … 100 mg/m³ | 5 months | IED 2010/75/EC | E50 |
NO | 0 … 250 mg/m³ | 12 months | IED 2010/75/EC | |
SO2 | 0 … 400 mg/m³ | 12 months | IED 2010/75/EC | |
N2O | 0 … 500 ppm | Kyoto protocol | ||
N2O | 0 … 50 mg/m³ | 6 months | IED 2010/75/EC |
Calibration interval (TÜV versions 7MB2355, 7MB2357, 7MB2358)
Component | Smallest measuring range | Calibration interval | Remarks | Z suffix |
---|---|---|---|---|
CO | 0 ... 200 mg/m³ | 1 month | 13th/27th BImSchV | TX3 |
NO | 0 ... 150 mg/m³ | 1 month | 13th/27th BImSchV | TX3 |
SO2 | 0 ... 400 mg/m³ | 1 month | 13th/27th BImSchV | TX3 |
AUTOCAL | AUTOCAL | Calibration with calibration gas | Comment (keep to technical specs) | |||||
Zero point | Calibration point | Zero point | Calibration point | Zero point | Calibration point | |||
Hours | Weeks | |||||||
IR components | 3 ... 24 | 3 ... 24 | o | 52 | ||||
O2 - electrical chemical sensor | Stable | 3 ... 24 | Stable | - | 52 | o | ||
O2 paramagnetic Cell | - | 3 ... 24 | x | x | 1 | o | at MB < 5% | |
- | 3 ... 24 | x | x | 8 | o | at MB > 5% | ||
O2 paramagnetic Cell | x | x | 3 ... 24 | - | o | 52 | at MB < 5% | |
x | x | 3 ... 24 | - | o | 52 | at MB > 5% | ||
H2S sensor | 3 | - | 3 | - | o | 12 | ||
o = with AUTOCAL, x = not applicable |
Calibration intervals, standard devices