19'''' - корпус Siemens

Чертеж
ULTRAMAT 6, 19“ unit, dimensions in mm
Схема подключения
Pin assignment (electrical and gas connections)
ULTRAMAT 6, 19“ unit, pin assignment
ULTRAMAT 6, 19“ unit, pin assignment of Autocal board and PROFIBUS connectors
ULTRAMAT 6, 19“ unit, gas and electrical connections (example: 2-channel version)
Технические данные
ULTRAMAT 6, 19“ unit | |
---|---|
General | |
Measuring ranges | 4, switchable internally and externally; autoranging is also possible |
Smallest possible measuring range | Depending on application, e.g., CO: 0 … 10 vpm, CO2: 0 … 5 vpm |
Largest possible measuring span | Depending on application |
Measuring range with suppressed zero | Every zero possible within 0 to 100 Vol.%, smallest possible measuring span 20% |
Characteristic | Linearized |
Position of use | Front panel vertical |
Conformity | CE identification EN 50081-1, EN 50082-2 |
Design, enclosure | |
Weight | Approx. 15 kg (with one IR channel), approx. 21 kg (with two IR channels) |
Degree of protection | IP20 according to EN 60529 |
Electrical characteristics | |
Electromagnetic compatibility (EMC) | According to standard requirements of NAMUR NE21 (08/98) |
Electrical safety | According to EN 61010-1, overvoltage category III |
Power supply | 100 to 120 V AC (rated range 90 to 132 V), 48 to 63 Hz or 200 to 240 V AC (rated range 180 to 264 V), 48 to 63 Hz |
Power consumption | 1-channel unit: approx. 40 VA 2-channel unit: approx. 70 VA |
Fuses | |
| 1T/250 (7MB2121), 1.6T/250 (7MB2123) |
| 0.63T/250 (7MB2121), 1T/250 (7MB2123) |
Gas inlet conditions | |
Perm. sample gas pressure | |
| |
| 600 to 1500 hPa (absolute) |
| 600 to 1300 hPa (absolute) |
| 600 to 1500 hPa (absolute) |
Sample gas flow | 18 to 90 l/h (0.3 to 1.5 l/min) |
Sample gas temperature | 0 to 50 °C |
Sample gas humidity | < 90 % RH (relative humidity) or depending on application, non condensing |
Time response | |
Warm-up period | With amb. temperature < 30 min (maximum accuracy achieved after 2 hours) |
Response time (T90 time) | Dependent on length of analyzer cell, sample gas line and damping |
Damping (electric time constant) | 0 to 100 s, programmable |
Dead time (purging time of gas path in analyzer at 1 l/min) | Approx. 0.5 to 5 s, depending on version |
Time for internal signal processing | < 1 s |
Pressure correction range | |
Pressure sensor | |
| 600 to 1200 hPa absolute |
| 600 to 1500 hPa absolute |
Measuring response (maximum accuracy achieved after 2 hours) | |
Output signal fluctuation | ± 0.1 % to ± 1 % of smallest possible measuring range specified on rating plate depending on application with the unit specific electronic time constant (corresponds to ± 0.33% at 2σ) |
Zero drift | < 1% of measuring range/week |
Measured-value drift | < 1% of measuring range/week |
Repeatability | ≤ 1% of respective measuring range |
Minimum detectable quantity | 1% of smallest measuring range |
Linearity error | < 0.5% of full-scale value |
Influencing variables (referred to 1000 hPa sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature) | |
Ambient temperature | < 1% of measuring range/10 K (for a constant temperature of the reception cell) |
Sample gas pressure | With pressure compensation: < 0.15% of span/1% change in atmospheric pressure without pressure compensation: < 1.5% of span/1% change in atmospheric pressure |
Sample gas flow | Negligible |
Power supply | < 0.1% of output signal span with rated voltage ± 10% |
Ambient conditions | Application-dependent influencing of measurement if ambient air contains measured component or cross-sensitive gases |
Electric inputs and outputs | |
Analog output | 0/2/4 to 20 mA. floating; load ≤ 750 Ω |
Relay outputs | 6, with changeover contacts, freely parameterizable, e.g. for range identification; loading capacity: 24 V AC/DC / 1 A floating, non sparking |
Analog inputs | 2, designed for 0/2/4 to 20 mA, for external pressure sensor and correction of influence of residual gas (correction of cross interference) |
Binary inputs | 6, designed for 24 V, floating, freely parameterizable, e.g. for range switching |
Serial interface | RS 485 |
Options | Autocal function with 8 additional binary inputs and 8 relay outputs, also with PROFIBUS PA and PROFIBUS DP |
Ambient conditions | |
Permissible ambient temperature | -30 to +70 °C during storage and transport, +5 to +45 °C during operation |
Permissible humidity | < 90 % RH (relative humidity) as annual average, during storage and transport (dew point must not be fallen below) |