ULTRAMAT/OXYMAT 6, 19“ unit, dimensions in mm
Схема подключения
Pin assignment (electrical and gas connections)
ULTRAMAT/OXYMAT 6, 19“ unit, pin assignment
ULTRAMAT/OXYMAT 6, 19“ unit, pin assignment of Autocal board and PROFIBUS connectors
ULTRAMAT/OXYMAT 6, 19“ unit, gas and electrical connections
Технические данные
General | |
---|---|
Position of use | Front panel vertical |
Conformity | CE identification EN 50081-1 and EN 50082-2 |
Design, enclosure | |
Weight | Approx. 21 kg |
Degree of protection | IP20 according to EN 60529 |
Electrical characteristics | |
EMC interference immunity | According to standard requirements of NAMUR NE21 (08/98) |
Electrical safety | According to EN 61010-1, overvoltage category III |
Power supply | 100 … 120 V AC (rated range 90 V … 132 V), 48 … 63 Hz or 200 … 240 V AC (rated range 180 V … 264 V), 48 … 63 Hz |
Power consumption | approx. 70 VA |
Fuses | 120 ... 120 V: F1/F2 = T 1.6 A; 200 ... 240 V: F1/F2 = T 1 A |
Electrical inputs and outputs (per channel) | |
Analog output | 0/2/4 … 20 mA, floating, max. load 750 Ώ |
Relay outputs | 6, with changeover contacts, freely selectable, e.g. for range identification; loading capacity: 24 V AC/DC/1 A, floating, non sparking |
Analog inputs | 2, designed for 0/2/4 … 20 mA, for external pressure sensor and correction of influence of residual gas (correction of cross interference) |
Binary inputs | 6, designed for 24 V, floating, freely selectable, e.g. for range switching |
Serial interface | RS 485 |
Options | Autocal function with 8 additional binary inputs and 8 relay outputs; also with PROFIBUS PA and PROFIBUS DP |
Ambient conditions | |
Perm. ambient temperature | -30 … +70 °C during storage and transport, +5 … +45 ºC during operation |
Permissible humidity | < 90% relative humidity, during storage and transport (dew point must not be fallen below) |
Technical data, ULTRAMAT channel | |
---|---|
Measuring ranges | 4, switchable internally and externally; autoranging is also possible |
Smallest possible measuring range | Depending on application, e.g.: CO: 0 … 10 vpm CO2: 0 … 5 vpm |
Largest possible measuring range | Depending on application |
Measuring ranges with suppressed zero | Any zero point is possible between 0 and 100%; smallest possible span 20% |
Characteristic | Linearized |
Gas inlet conditions | |
Permissible sample gas pressure | |
| 600 … 1500 hPa (absolute) |
| 600 … 1300 hPa (absolute) |
Sample gas flow | 18 … 90 l/h (0.3 … 1.5 l/min) |
Sample gas temperature | 0 … 50 ºC |
Sample gas humidity | < 90 % relative humidity or depending on application, without condensation |
Time response | |
Warm-up period | With amb. temperature < 30 min (maximum accuracy achieved after 2 hours) |
Response time (T90 time) | Dependent on length of analyzer cell, sample gas line and parameterizable damping |
Damping (electric time constant) | 0 to 100 s, programmable |
Dead time (purging time of gas path in analyzer at 1 l/min) | Approx. 0.5 to 5 s, depending on version |
Time for internal signal processing | < 1 s |
Pressure correction range | |
Pressure sensor | |
| 600 … 1200 hPa absolute |
| 600 … 1500 hPa absolute |
Measuring response (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature) | |
Output signal fluctuation | ± 0,1% ... ± 1% of smallest possible measuring range specified on rating plate depending on the unit electronic time constant (corresponds to ± 0.33% with 2σ) |
Zero drift | < 1% of measuring range/week |
Measured-value drift | < 1% of measuring range/week |
Repeatability | ≤ 1% of respective measuring range |
Linearity error | < 0.5% of full-scale value |
Influencing variables (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature) | |
Ambient temperature | < 1% of measuring range/10 K (only with stable EK temperature) |
Sample gas pressure | With pressure compensation: < 0.15% of span/1% change in atmospheric pressure without pressure compensation: < 1.5% of span/1% change in atmospheric pressure |
Sample gas flow | Negligible |
Power supply | < 0.1% of output signal span with rated voltage ± 10% |
Ambient conditions | Application-dependent influencing of measurement if ambient air contains measured component or cross-sensitive gases |
Technical data, OXYMAT channel | |
---|---|
Measuring ranges | 4, switchable internally and externally; autoranging is also possible |
Smallest possible measuring span (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature) | 0.5% v/v, 2% v/v or 5% v/v O2 |
Largest possible measuring span | 100 % v/v O2 |
Measuring ranges with suppressed zero | Any zero point is possible between 0 and 100% v/v as long as a suitable reference gas is used |
Gas inlet conditions | |
Permissible sample gas pressure | |
| 500 … 3000 hPa absolute |
| |
| 500 … 1500 hPa absolute |
| 500 … 1300 hPa absolute |
Sample gas flow | 18 … 60 l/h (0.3 … 1 l/min) |
Sample gas temperature | 0 … 50 ºC |
Sample gas humidity | < 90% RH (relative humidity) |
Time response | |
Warm-up period | With ambient temperature < 30 min (maximum accuracy achieved after 2 hours) |
Reading delay time | min. 1.5 … 3.5 s, depending on version |
Damping (electric time constant) | 0 to 100 s, programmable |
Dead time (purging time of gas path in analyzer at 1 l/min) | Approx. 0.5 … 2.5 s depending on version |
Time for internal signal processing | < 1 s |
Pressure correction range | |
Pressure sensor | |
| 500 … 2000 hPa absolute |
| 500 … 3000 hPa absolute |
Measuring response (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature) | |
Output signal fluctuation | < 0.75% of smallest possible measuring range specified on rating plate with an electronic time constant of 1 s (corresponds to ± 0.25% with 2σ) |
Zero drift | < 0.5%/month of smallest possible meas. span specified on rating plate |
Measured-value drift | < 0.5%/month of respective measuring span |
Repeatability | < 1%/month of respective measuring span |
Linearity error | < 1%/month of respective measuring span |
Influencing variables (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature) | |
Ambient temperature | < 0.5%/10 K referred to the smallest possible measuring span according to rating plate with span 0.5%: 1%/10 K |
Sample gas pressure (with air (100 hPa) as reference gas, a correction of the atmospheric pressure fluctuations is only possible when the sample gas is vented to ambient air) | Without pressure compensation: < 2% of measuring span/1% change in pressure; with pressure compensation: < 0.2% of measuring span/1% change in pressure |
Residual gases | Deviation in zero point corresponding to paramagnetic or diamagnetic deviation of residual gas |
Sample gas flow | < 1% of smallest possible measuring span according to rating plate with a change in flow of 0.1 l/min within the permissible flow range |
Power supply | < 0.1% of output signal span with rated voltage ± 10% |