19 unit Siemens

Чертеж
OXYMAT 64, 19" unit, size in mm
Схема подключения
Pin assignment (electrical connections)
OXYMAT 64, 19" unit, pin assignment
OXYMAT 64, 19" unit, pin assignment of the AUTOCAL plate and PROFIBUS plug
Gas connections and pin assignment
OXYMAT 64, 19" unit, gas connections and electrical connections
Технические данные
General | |
---|---|
Measurement ranges | 4, internally and externally switchable; automatic measurement range changeover also possible |
Smallest possible measuring span (relating to sample gas pressure 1000 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature) | 0 ... 10 vpm O2 |
Largest possible measuring span | 0 ... 99999 vpm |
Operating position | Front wall vertical |
Conformity | CE mark in accordance with EN 50081-1, EN 50082-2 and RoHS |
Design, enclosure | |
Degree of protection | IP20 according to EN 60529 |
Weight | Approx. 11 kg |
Electrical characteristics | |
EMC (Electromagnetic Compatibility) | In accordance with standard requirements of NAMUR NE21 (08/98) and EN 61326 |
Electrical safety | According to EN 61010-1, overvoltage category II |
Auxiliary power | 100 ... 120 V AC (nominal use range 90 ... 132 V), 48 ... 63 Hz or 200 ... 240 V AC (nominal use range 180 ... 264 V), 48 ... 63 Hz |
Power consumption | Approx. 37 VA |
Fuse values | 100 ... 120 V: 1.0T/250 200 ... 240 V: 0.63T/250 |
Gas inlet conditions | |
Sample gas flow | |
| 7.5 l/h |
| 15 ... 30 l/h |
Permissible sample gas pressure | |
| 2000 hPa (abs.) |
| 2000 ... 6000 hPa (abs.) |
Sample gas temperature | 0 ... 50 °C |
Sample gas humidity | < 1% relative humidity |
Dynamic response | |
Heating time | at room temperature < 30 min (the technical specification is will be observed after 2 hours) |
Display delay (T90) | ~ 1 min upon switch from N2 to air |
Damping (electrical time constant) | 0 ... 100 sec, parameterizable |
Dead time (purge time of the gas route in the device at 125 ml/min) | 10 ... 30 sec |
Time for device-internal signal processing | < 1 sec |
Pressure correction range | |
Pressure sensor internal | 800 ... 1100 hPa (abs.) (permissible sample gas pressure, see gas inlet conditions) |
Measuring behavior (relating to sample gas pressure 1000 hPa absolute, 7.5 l/min sample gas flow and 25 °C ambient temperature) | |
Output signal fluctuation | < 1 % of the smallest possible measurement range as per nameplate with electronic damping constant of 1 s |
Zero point drift | < ± 1% of the current measuring span/month |
Measured value drift | < ± 1% of the current measuring span/month |
Repeat precision | < 3% of the current measuring span |
Minimum detectable quantity | 1% of current measurement range, < 0.1 vpm in measurement range 0 ... 10 vpm |
Deviation of linearity | < 2% of the current measuring span |
Influencing variables (relating to sample gas pressure 1000 hPa absolute, 7.5 l/min sample gas flow and 25 °C ambient temperature) | |
Ambient temperature | < 2%/10 K relating to the current measuring span |
Sample gas pressure only possible if the sample gas can flow out into the ambient air | When pressure compensation has been switched off: < 1% of the current measuring span/1% pressure change When pressure compensation has been switched on: < 0.2% of the current measuring span/1% pressure change |
Carrier gases, zero point deviation | |
| Only gases with non-combustible carrier gas components can be introduced |
| Carrier gas concentration of 10 vpm H2; CO and CH4 have a low diagonal effect; higher CVs (concentration values) are negligible |
Sample gas flow | < 2% of the smallest possible measuring span with a flow change of 10 ml/min |
Auxiliary power | < 0.1% of the current measurement range with nominal voltage ± 10% |
Electrical inputs and outputs | |
Analog output | 0/2/4 ... 20 mA, 4 ... 20 mA (NAMUR), potential-free; apparent ohmic resistance max. 750 Ω |
Relay outputs | 6, with changeover contacts, freely parameterizable, e.g. for measurement range identification; load capacity: 24 V AC/DC / 1 A, potential-free |
Analog inputs | 2, designed for 0/2/4 ... 20 mA for pressure sensor external and carrier gas inflow correction (diagonal gas correction) |
Binary inputs | 6, designed for 24 V, potential-free, freely parameterizable, e.g. for measurement range changeover |
Serial interface | RS 485 |
Options | AUTOCAL function each with 8 additional binary inputs and relay outputs, also with PROFIBUS PA or PROFIBUS DP |
Climatic conditions | |
Permissible ambient temperature | -40 ... +70 °C during storage and transportation, 5 ... +45 °C during operation |
Permissible humidity | < 90% relative humidity within average annual value, during storage and transportation (no passing below the dew point) |