19'''' - корпус Siemens

Чертеж
FIDAMAT 6, 19“ unit, dimensions in mm
Схема подключения
Pin assignment (electrical and gas connections)
FIDAMAT 6, 19“ unit, pin assignment
FIDAMAT 6, 19“ unit, pin assignment of the Autocal board and PROFIBUS connectors
FIDAMAT 6, gas and electrical connections, version with pump
Технические данные
General | |
---|---|
Measuring ranges | 4, switchable between internal and external; manual and automatic measuring range switchover possible |
Smallest possible measuring span | 0 ... 10 ppm |
Verification limit | 0.1 ppm |
Largest possible measuring span | 99.999 ppm |
Concentration units | ppm, C1, C3, C6 or mgC/m3 |
Automatic measuring range switchover | Hysteresis, selectable |
Measured value display | Digital concentration display (5 digits with floating point) |
Resolution of the digital display | 0.1% of the measured value |
Position of use | Front panel vertical |
Conformity | CE mark |
Oven temperature | Adjustable 100 ... 200 °C |
Design, enclosure | |
Degree of protection | IP20 in accordance with EN 60529 |
Weight | approx. 23 kg |
Electrical characteristics | |
Auxiliary power | 100 ... 120 V AC (rated operational range 90 V ... 132 V), 48 ... 63 Hz or 200 ... 240 V AC (rated operational range 180 V ... 264 V), 48 ... 63 Hz |
Power consumption | approx. 150 VA in operation, |
EMC (Electromagnetic Compatibility) | in accordance with standard NAMUR NE21 requirements (08/98) |
Electrical safety | in accordance with EN 61010-1, overvoltage category II |
Fuse values | 100 ... 120 V: 4.0T/250 |
Gas inlet conditions | |
Permissible sample gas pressure | |
| < 2000 hPa abs. |
| 600 … 1100 hPa |
Sample gas flow | 18 ... 60 l/h (0.3 ... 1 l/min) |
Sample gas temperature | 0 ... 200 °C |
Sample gas humidity | < 90 % RH |
Time response | |
Warm-up period | approx. 2-3 h at room temperature |
Response time (T90) | 2 ... 3 s |
Attenuation (electrical time constant) | 0 ... 100 s, parameterizable |
Dead time (purging time of the gas line in the device at 1 l/min) | with filter 2 ... 3 s |
Time for signal processing in the device | < 1 s |
Measuring response | |
Output signal fluctuation | < 0.75% of the smallest possible measuring range according to the rating plate at an electronic attenuation constant of 1 s (this corresponds to ± 0.25% at 2σ) |
Zero drift | < 0.5%/month of the smallest possible measuring span according to the rating plate |
Measured value drift | < 1%/week of the appropriate measuring span |
Repetition precision | < 1% of the appropriate measuring span |
Minimum detectable quantity | 0.1 ppm (version for ultra-pure gas measurement: 50 ppb) |
Linearity error | < 1% of the appropriate measuring span |
Influencing variables | |
Ambient temperature | < 1%/10 K related to the least possible measuring span according to the rating plate |
Ambient pressure | < 1%/50 hPa |
Sample gas pressure | < 2% of the measuring span/1% pressure change (within 600 … 1100 hPa) |
Auxiliary power | < 1% at a change in the output signal span of ± 10% |
Frequency | < 1% |
Position influence | < 1% at an inclination < 15° |
Electrical inputs and outputs | |
Analog output | 0/2/4 ... 20 mA, floating; max. load 750 Ω |
Relay outputs | 6, with freely parameterizable changeover contacts for identifying the measuring range, for example; load rating: 24 V AC/DC /1 A, floating |
Analog inputs | 2, designed for 0/2/4 to 20 mA for pressure pick-ups externally and for correcting the influence of residual gas (interfering gas) |
Binary inputs | 6, designed for 24 V, floating, freely parameterizable, for measuring range switchover, for example |
Serial interface | RS 485 |
Options | Autocal function with 8 additional binary inputs and relay outputs, also with PROFIBUS PA or PROFIBUS DP |
Climatic conditions | |
Permissible ambient temperature | 5 ... +45 °C in operation, -30 ... +70 °C for storage and transport |
Permissible humidity | < 90% RH (RH: relative humidity) mid-year, for storage and transport (dew point must not be fallen below) |
FIDAMAT 6 with pump and heated oven, with combustion air connection | |||||
---|---|---|---|---|---|
Gases | Operating pressure | ||||
Input pressure | Pump startup | Flow through FID | Flow through bypass | ||
without | with | ||||
hPa (abs.) | hPa (abs.) | hPa (abs.) | ml/min | ml/min | |
Combustion gas | 3000 ... 5000 | 1000 ± 20 | ~ 25 | — | |
Combustion air | 3000 ... 5000 | 1420 ± 20 | 1500 | ~ 320 | ~ 500 |
Sample gas | ~ 1000 | — | 1500 ± 2 | ~ 3 | ~ 1000 |
Zero gas | 3500 ... 4000 | — | 1500 ± 2 | ~ 3 | ~ 1000 |
Calibration gas | 3500 ... 4000 | — | 1500 ± 2 | ~ 3 | ~ 1000 |
FIDAMAT 6 without pump and with heated oven, with combustion air connection | |||||
---|---|---|---|---|---|
Gases | Operating pressure | ||||
Input pressure | Sample/calibration gas | Flow through FID | Flow through bypass | ||
without | with | ||||
hPa (abs.) | hPa (abs.) | hPa (abs.) | ml/min | ml/min | |
Combustion gas | 3000 ... 5000 | 1000 ± 20 | ~ 25 | — | |
Combustion air | 3000 ... 5000 | 1485 ± 2 | 1500 | ~ 320 | ~ 300 |
Sample gas | ~ 2000 | — | 1500 ± 2 | ~ 3 | ~ 500 |
Zero gas | ~ 2000 | — | 1500 ± 2 | ~ 3 | ~ 500 |
Calibration gas | ~ 2000 | — | 1500 ± 2 | ~ 3 | ~ 500 |
FIDAMAT 6 with pump and with heated oven, ambient air = combustion air | |||||
---|---|---|---|---|---|
Gases | Input pressure | Operating pressure | Flow through FID | Flow through bypass | |
Pump startup | |||||
without | with | ||||
hPa (abs.) | hPa (abs.) | hPa (abs.) | ml/min | ml/min | |
Combustion gas | 3000 ... 5000 | 1000 ± 20 | ~ 25 | xxxx | |
Sample gas | ~ 1000 | Xxxx | 1500 ± 2 | ~ 3 | ~ 500 |
Zero gas | 3500 ... 4000 | Xxxx | 1500 ± 2 | ~ 3 | ~ 500 |
Calibration gas | 3500 ... 4000 | xxxx | 1500 ± 2 | ~ 3 | ~ 500 |