Set CEM 1 Siemens

Область применения



The monitoring of emission components in flue gases is one of the most important topics for continuous gas analysis. This is as a result of legislation for monitoring emissions, e.g. for large combustion plants, and also due to the requirements of companies operating process plants who can draw conclusions on the process efficiency from the gas analyses, e.g. with boiler control, DENOX and DESOX plants.

The market requires a reliable complete system which is specially designed for the application. The Set CEM 1 (Continuous Emission Monitoring) offered by Siemens is a system which reliably covers all requirements associated with sampling, sample preparation, and gas analysis.

It is possible to determine the concentrations of the gaseous components CO, CO2, NO, NOx, SO2, O2, HCl, HF, NH3 and H2O.

The ULTRAMAT 23 and OXYMAT 6 are used for the extractive, continuous gas analysis.

The standardized Set CEM provides great clarity and simple configuration facilities. Different versions mean that it is possible to appropriately adapt the system to the requirements. Standardization also means that not all imaginable versions can be included, and that it may not be possible to implement special requirements such as armored cables, customer-specific documentation or specific conductor labeling without an extra charge.

Обзор



The Set CEM is a standardized system specially for monitoring the emission components in flue gases.

Дизайн



Starting with a mounting frame with sample preparation system, it is possible to add additional units as options. These include:

  • Sampling probe with weather protection hood
  • Heated sample gas line
  • Analyzers
  • Air-conditioning unit
  • NO2/NO converter
  • Sample preparation extension for an additional ULTRAMAT 23 analyzer
  • Single and dual (electrically isolated, not electrically isolated) analog signal processing
  • Power supply modules (115 V, 230 V, 400 V)
  • Outer panels with steel-plate door or with window
  • Single-pole and double-pole fusing
  • Condensation bottle
  • Coalescence filter

Sampling probe

The standard probe is fitted with a DIN flange DN 65, PN 6. The probe is provided with a self-regulating heater, and has a power consumption of 400 VA. It is supplied with a weather protection hood and 2 µm filter. The maximum dust concentration at the sampling point should not exceed 2 g/m³. The sampling pipe is 1 000 mm long, is made of stainless steel, and has dimensions of 20 x 1.5 mm. The sample gas temperature must not exceed 600 °C.

It is also possible to purchase the Set CEM 1 without sample probe.

Heated sample gas line

The temperature of the heated line is regulated at 200 °C. The power consumption is 100 VA per meter. The internal core is made of PTFE 4/6. The minimum length of the heated line is 5 m, and it can be extended up to 35 m. Lengths greater than 35 m can be provided upon special request. If desired, the system can also be supplied without a heated sample gas line.

Mounting frame

The basis of each Set CEM is the mounting frame with hinged frame (40 HU) for installation of up to five 19" rack units. The mounting frame includes a standardized sample preparation device designed for an ULTRAMAT 23.

The sample preparation system includes a 3/2-way solenoid valve, 3-way switchover ball valve, regulating valve, corrosion-resistant sample gas pump (power consumption 60 VA), condensation trap, room air suction filter with filter element, LOGO for digital display of individual signals in the cabinet, 24 V DC power supply unit (power consumption 70 VA) and moisture sensor with flow cell. Also included is a sample gas cooler (power consumption 200 VA) with integral preliminary condensation trap, heat exchanger, hose pump and teflon filter.

The external dimensions without plinth are 2 000 x 800 x 800 mm (H x W x D). A cabinet depth of 600 mm is also available (not suitable for LDS 6). Hoses and cables can be connected from the left or right. A distance of 500 mm must be provided on the left or right at the site for introduction of the hoses and cables.

In addition to the sheet-steel mounting frames for indoor installation, an FRP version is also available for outdoor use. The FRP cabinet is always provided complete with side panels and plinth. The external dimensions are 2 080 x 800 x 600 mm (H x W x D). The FRP cabinet cannot be combined with the LDS 6.

Side panels with doors

Optional outer panels can be selected for the sheet-steel mounting frames. This possibility allows use of the Set CEM in analysis cabinets as a rack design, or in halls as a cabinet design, where a protection class of max. IP54 is required in the latter case. Either a sheet-steel door without window or a glass door can be selected.

Plinth

Plinths with a height of 100 and 200 mm are additionally available.

Cabinet cooling and ventilation

Optionally available are a fan with outlet filter, an air-conditioning unit for indoor installation, and an air-conditioning unit for outdoor installation. The system can be ordered without a fan or air-conditioning unit if the side panels and the door with window are omitted.

The fan with outlet filter has a power consumption of 60 VA, and is fitted in the cabinet wall. The delivery also includes a thermostat with a power consumption of 25 VA.

The air-conditioning unit has a cooling power of 820 VA.

Frost protection heater

The power consumption of the optional cabinet heater is 500 VA. The delivery includes a thermostat with a power consumption of 25 VA for controlling the frost protection heater.

If desired, the system can also be supplied without a frost protection heater.

Fusing of the analog signals

In addition to single-pole fusing of the electronic consumers, it is possible to provide double-pole fusing.

Removal of condensation

Either a 5 or 10 liter condensation bottle can be provided as an option. It is also possible to order the system without a condensation bottle if the condensation can be removed on-site.

NO2/NO converter

The mounting frame and cabinets can be optionally extended by a 19" rack unit with NO2/NO converter with carbon cartridge. The power consumption is 520 VA. The flow rate is 90 l/min. A NO2/NO converter is required if the share of NO2 in the total NOx is greater than 5% and/or if total NOx is to be always determined.

Power supply

The system can be designed either for 115 V AC, 230 V AC or 400 V AC (-15%, +10%) with 50 or 60 Hz.

Analog signal processing

As standard, the analog signals are simply connected to isolating terminals. As an option, the analog signals can be processed twice without electrical isolation by a diode module, or twice with electrical isolation.

Analyzers

The standardized set is prepared for an ULTRAMAT 23. The system can be supplemented by a second ULTRAMAT 23, OXYMAT 6 and/or LDS 6. Various measured components and measuring ranges can be selected. Other combinations of components and ranges are available on request. In such a case it must be checked whether the desired certificates and approvals are available. The analyzers used, the measured components and the measuring ranges are described briefly below.

Details on the analyzers, alternative components and ranges can be found under the topics "Continuous gas analyzers, extractive" and "Continuous gas analyzers, in-situ".

ULTRAMAT 23: CO, NO

For measuring two infrared components.

Component

Smallest tested measuring range

Switchable to

CO

0 ... 150 mg/Nm³

0 ... 750 mg/Nm³

NO

0 ... 100 mg/Nm³

0 ... 500 mg/Nm³



One or two measuring ranges can be freely set within the limits. The ULTRAMAT 23 carries out automatic self-calibration with ambient air. The power consumption is 60 VA.

ULTRAMAT 23: CO, NO, SO2

For measuring three infrared components.

Component

Smallest tested measuring range

Switchable to

CO

0 ... 250 mg/Nm³

0 ... 1 250 mg/Nm³

NO

0 ... 400 mg/Nm³

0 ... 2 000 mg/Nm³

SO2

0 ... 400 mg/Nm³

0 ... 2 000 mg/Nm³



One or two measuring ranges can be freely set within the limits. The ULTRAMAT 23 carries out automatic self-calibration with ambient air. The power consumption is 60 VA.

ULTRAMAT 23: CO, NO, CO2

For measuring three infrared components.

Component

Smallest tested measuring range

Switchable to

CO

0 ... 250 mg/Nm³

0 ... 1 250 mg/Nm³

NO

0 ... 400 mg/Nm³

0 ... 2 000 mg/Nm³

CO2

0 ... 5%

0 ... 25%



One or two measuring ranges can be freely set within the limits. The ULTRAMAT 23 carries out automatic self-calibration with ambient air. The power consumption is 60 VA.

The component CO2 has not been type approved by the TÜV.

ULTRAMAT 23: CO2

For measuring one infrared component.

CO2: smallest measuring range: 0 to 5%, largest measuring range: 0 to 25%.

One or two limits can be freely set within the limits. The ULTRAMAT 23 carries out automatic self-calibration with ambient air. The power consumption is 60 VA.

The component CO2 has not been type approved by the TÜV.

The ULTRAMAT 23 analyzers can be optionally equipped with an electrochemical oxygen sensor.

O2: tested measuring ranges 0 to 10/25%

OXYMAT 6: O2

For paramagnetic measurement of oxygen. Instead of ULTRAMAT 23 with electrochemical cell.

O2: tested measuring ranges 0 to 10% / 0 to 25%

Sample chamber without flow-type compensation branch, made of stainless steel 1.4571.

Preparation of sample preparation system for second ULTRAMAT 23

The standard system with sample preparation system and electronics is prepared for one ULTRAMAT 23. If a second ULTRAMAT 23 is to be fitted, this option must be selected so that the sample preparation system and electronics are extended accordingly.

Additional filter

In addition to the fine filter and moisture filter which are always present, a coalescence filter can be optionally fitted in the sample preparation device.

LDS 6: HCl

Component

Smallest tested measuring range

HCl

0 ... 15 mg/Nm³

  


Application for channel 1: Emission monitoring

The power consumption is 50 VA. Suitable for connection of non-Ex sensors, including non-Ex-protected sensor electronics.

The delivery includes a pair of sensors for instrument air or N2 on the process side. The pair of sensors is designed for a moderate flow rate of 0 to 120 l/min. The 400 mm long purging tubes are made of stainless steel. The process connection is DN 65, PN 6. The power consumption is 2 VA.

Limitation:
Applies to measurement paths > 2 000 mm, applies to gases with a methane content < 15 mg/m³. Necessary gas temperature between 120 and 210 °C.

LDS 6: HCl / H2O

Component

Smallest tested measuring range

HCl

0 ... 15 mg/Nm³

H2O

0 ... 30 %



Application for channel 1: Emission monitoring

The power consumption is 50 VA. Suitable for connection of non-Ex sensors, including non-Ex-protected sensor electronics.

The delivery includes a pair of sensors for instrument air or N2 on the process side. The pair of sensors is designed for a moderate flow rate of 0 to 120 l/min. The 400 mm long purging tubes are made of stainless steel. The process connection is DN 65, PN 6. The power consumption is 2 VA.

Limitation:
Applies to measurement paths > 2 000 mm, applies to gases with a methane content < 15 mg/m³. Necessary gas temperature between 120 and 210 °C.

LDS 6: HF

HF: Smallest possible measuring range depends on the gas composition.

Application for channel 1: Emission monitoring

The power consumption is 50 VA. Suitable for connection of non-Ex sensors, including non-Ex-protected sensor electronics.

The delivery includes a pair of sensors for instrument air or N2 on the process side. The pair of sensors is designed for a moderate flow rate of 0 to 120 l/min. The 400 mm long purging tubes are made of stainless steel. The process connection is DN 65, PN 6. The power consumption is 2 VA. The HF measurement has not been type approved by the TÜV.

Limitation:
Component has not been type approved by the TÜV. Necessary gas temperature between 0 and 150 °C.

LDS 6: HF/H2O

HF: Smallest possible measuring range depends on the gas composition.

H2O: Smallest tested measuring range: 0 to 30%

Application for channel 1: Emission monitoring

The power consumption is 50 VA. Suitable for connection of non-Ex sensors, including non-Ex-protected sensor electronics.

The delivery includes a pair of sensors for instrument air or N2 on the process side. The pair of sensors is designed for a moderate flow rate of 0 to 120 l/min. The 400 mm long purging tubes are made of stainless steel. The process connection is DN 65, PN 6. The power consumption is 2 VA. The HF measurement has not been type approved by the TÜV.

Limitation:
Component has not been type approved by the TÜV. Necessary gas temperature between 0 and 150 °C.

LDS 6: NH3

Component

Smallest tested measuring range

NH3

0 ... 20 mg/Nm³

  


Application for channel 1: Emission monitoring

The power consumption is 50 VA. Suitable for connection of non-Ex sensors, including non-Ex-protected sensor electronics.

The delivery includes a pair of sensors for instrument air or N2 on the process side. The pair of sensors is designed for a moderate flow rate of 0 to 120 l/min. The 400 mm long purging tubes are made of stainless steel. The process connection is DN 65, PN 6. The power consumption is 2 VA.

Limitation:
Applies to measurement paths > 1 250 mm. Necessary gas temperature between 0 and 150 °C.

LDS 6: NH3/ H2O

Component

Smallest tested measuring range

NH3

0 ... 20 mg/Nm³

H2O

0 ... 15 %



Application for channel 1: Emission monitoring

The power consumption is 50 VA. Suitable for connection of non-Ex sensors, including non-Ex-protected sensor electronics.

The delivery includes a pair of sensors for instrument air or N2 on the process side. The pair of sensors is designed for a moderate flow rate of 0 to 120 l/min. The 400 mm long purging tubes are made of stainless steel. The process connection is DN 65, PN 6. The power consumption is 2 VA.

Limitation:
Applies to measurement paths > 1 250 mm. Necessary gas temperature between 0 and 150 °C.

Hybrid cable

A hybrid cable is required to connect a central unit to one pair of sensors. Versions for 5, 10, 25, 40 and 50 m are available. Cable lengths cannot be combined. Lengths greater than 50 m can be ordered on request.

Sensor cable

A sensor cable is required to connect one pair of sensors. Versions for 5, 10 and 25 m are available. Cable lengths cannot be combined. Lengths greater than 25 m can be ordered on request.

Electrical preparation for dust measurement

Electrical preparation for connection of an external dust measurement to the system.

Electrical preparation for flow measurement

Electrical preparation for connection of an external flow measurement to the system.

Electrical preparation for pressure measurement

Electrical preparation for connection of an external pressure measurement to the system.

Electrical preparation for temperature measurement

Electrical preparation for connection of an external temperature measurement to the system.

Electrical preparation for emission data memory on rail module

Electrical preparation for connection of an emission data memory on rail module to the system.

Electrical preparation for emission data memory in 19" rack unit

Electrical preparation for connection of an emission data memory in 19" rack unit to the system.

Additional LOGO for four or more 19" rack units

Sets with more than three 19" rack units integrated require a LOGO extension module. The delivery also includes connection and programming.

Core end labeling

It is optionally possible to order core end labeling according to the Siemens standard (VDE 0100 Part 200).

Documentation

The Siemens standard documentation is available in German, English or French.

The documentation includes gas path diagram, circuit diagram, terminal diagram, installation diagram, consumable materials list, signal list, cable list and parts list. Also included are technical data sheets and manuals for the components and devices used. The documentation language for parts provided by other suppliers may deviate. Plant description, LOGO program and test certificates are also included in the documentation.

Example setup, may deviate.

Функции



A sample is extracted via the thermostatically-controlled sample gas probe. The dust concentration may be up to 2 g/m³, the sample gas temperature up to 600 °C. The gas is transported to the analysis cabinet via a heated sample gas line. The heating prevents condensate. The gas cooler cools and dries the sample in the analysis cabinet. Condensate is drained. The level in the condensate trap is monitored. For safety purposes, a coalescence filter can be provided in addition to the fine filter and moisture filter which are always present. The sample gas is analyzed by analyzers such as the ULTRAMAT 23, OXYMAT 6 and LDS 6. The ULTRAMAT 23 operates on the basis of molecular-specific absorption of infrared radiation or with an electrochemical oxygen measuring cell. The OXYMAT 6 is an analyzer for paramagnetic oxygen measurements. The in-situ LDS 6 laser diode spectrometer operates according to the molecular-specific absorption of near-IR radiation. The delivery may also include an NO2/NO converter which permits measurement of total nitrogen oxides. In order to qualify the set for low or high temperature ranges (-5, +45 °C), it is possible to use a cabinet heater or air-conditioning unit. Power supply versions are available for 115, 230 or 400 V AC. Electronic consumers can be provided with single-pole or double-pole fusing. The components of the sample preparation system and the analyzers are connected to LOGO modules via a binary signal, and transmit maintenance requirements. The analog signals can be processed either singly or twice. Electrical isolation is additionally possible for the double processing.

Чертеж



Example setup, may deviate, dimensions in mm

Особенности



Standardized complete system

  • Highly exact and reliable monitoring of emissions in flue gases. System-specific QAL 1 certificate
  • Modular complete package with gas sampling system, sample gas preparation device and gas analyzers from one source
  • Simple and fast to configure
  • Tried and tested, harmonized and reliable set
  • Low costs for purchasing and operation

Proven, suitability-tested technologies

  • Continuous determination of up to 8 measured components
  • In-situ measurements without sampling and preparation, using LDS 6 laser diode spectrometer
  • Use of ULTRAMAT 23 with cleanable cells and automatic calibration with ambient air as well as optional electrochemical oxygen measurement
  • Paramagnetic oxygen measurement with OXYMAT 6

Simple operation

  • Intuitive operation
  • Configuration on large displays using plain text, in several languages

Simple maintenance

  • Maintenance-friendly cabinet design with hinged frame and uniform design
  • Digital display of maintenance requests on LOGO modules

Технические данные



Climatic conditions

 

Ambient temperature

0 ... 35 °C

  • With heater in sheet-steel cabinet

Max. -5 °C

  • With heater in FRP cabinet

Max. -15 °C

  • With air-conditioning

Max. 45 °C

Relative humidity

70 %, non-condensing

Corrosive atmosphere

No

Gas inlet conditions

 

Max. sample gas pressure at inlet to sample preparation system

0.5 barg

Min. sample gas temperature at inlet to sample preparation system

180 °C

Dust content at inlet to sample preparation system

Dust-free

Sampling probe

Sampling tube 20 x 1.5,
1 000 mm long, stainless steel,
flange: DN 65, PN 6

Max. sample gas pressure at sampling probe

0.5 barg

Max. sample gas temperature at sampling probe

600 °C

Max. dust content at sampling probe

2 g/Nm³

Sample gas must not be flammable or explosive.

Power supply

 

Supply 1

115 V AC (-15%, +10%)

Supply 2

230 V AC (-15%, +10%)

Supply 3

400 V AC (-15%, +10%)

Connections

 

Hose material

Teflon

Cables

Not armored, not halogen-free

Electrical design

According to IEC

Cable ID

Individual core labeling as option

Fusing of electronic consumers

1-pole; 2-pole as option

Duplication of analog signals

Not electrically isolated as option

Electrically isolated as option

Installation

 

Site

 
  • In sheet-steel cabinet/frame

Indoor installation

  • In FRP cabinet

Outdoor installation

Ex zone

Non-Ex area

System design

 

Version

Mounting frame or cabinet

Cabinet degree of protection

IP54

Automatic calibration

Yes, with ULTRAMAT 23

Detailed information on the analyzers

Process analytics, topics "Continuous gas analyzers, extractive" ULTRAMAT 23, OXYMAT 6 and "Continuous gas analyzers, in-situ" LDS 6

Dimensions (without plinth)

 

Depth of sheet-steel frame

 
  • 800 mm (without plinth)

2 000 x 800 x 800 mm (H x W x D)

  • 600 mm (without plinth)

2 000 x 800 x 600 mm (H x W x D)

FRP cabinet (with plinth)

2 080 x 900 x 600 mm (H x W x D)

It is necessary to allow a space of 500 mm on the left or right for introduction of the cables.